{ "cent-01_calibration_sec_0": "# CENT-01 Laboratory Centrifuge Calibration and Operation Manual", "cent-01_calibration_sec_1": "## 1. Introduction\nThe CENT-01 is a high-speed, precision laboratory centrifuge designed for biological sample separation. This manual covers the standard operating procedures, calibration protocols, and safety limits for the CENT-01 model.", "cent-01_calibration_sec_2": "## 2. General Operation\nTo ensure safety and equipment longevity, operators must adhere to strict payload-to-RPM ratios. Excessive RPM on heavy payloads can cause severe resonant vibration, potentially leading to catastrophic hardware failure or sample destruction. \nThe device supports a maximum theoretical rotational speed of 15,000 RPM under zero-load conditions.", "cent-01_calibration_sec_3": "## 3. Calibration Tables\nThe following tables outline the maximum safe RPM and expected vibration (RMS g-force) limits based on the payload mass. \n\n### 3.1. Standard Rotor Calibration\n\n| Payload Mass (g) | Max Safe RPM | Warning Vibration Limit (g) | Critical Shutdown Limit (g) |\n|------------------|--------------|-----------------------------|-----------------------------|\n| 10g | 12,000 RPM | 0.25 g | 0.50 g |\n| 50g | 8,000 RPM | 0.30 g | 0.60 g |\n| 100g | 5,000 RPM | 0.35 g | 0.70 g |\n| 250g | 3,000 RPM | 0.40 g | 0.80 g |\n| 500g | 1,500 RPM | 0.50 g | 1.00 g |\n\n### 3.2. Heavy Duty Rotor Calibration\n\n| Payload Mass (g) | Max Safe RPM | Warning Vibration Limit (g) | Critical Shutdown Limit (g) |\n|------------------|--------------|-----------------------------|-----------------------------|\n| 100g | 7,000 RPM | 0.30 g | 0.65 g |\n| 250g | 4,500 RPM | 0.35 g | 0.75 g |\n| 500g | 2,500 RPM | 0.40 g | 0.85 g |\n| 1000g | 1,000 RPM | 0.55 g | 1.10 g |", "cent-01_calibration_sec_4": "## 4. Emergency Procedures\nIf the telemetry system detects a vibration exceeding the Critical Shutdown Limit, the centrifuge will automatically engage the EMERGENCY_STOP protocol.\n- **Rule 4.A**: Under no circumstances should an operator attempt to override an EMERGENCY_STOP.\n- **Rule 4.B**: After an EMERGENCY_STOP, the centrifuge requires a full mechanical reset and recalibration before the next run.\n- **Rule 4.C**: Resonance frequencies typically occur at 7,200 RPM. Avoid prolonged operation in the 7,100-7,300 RPM band.", "cent-01_calibration_sec_5": "## 5. Maintenance\nRotor inspection must be conducted every 500 operational hours. Lubrication of the main spindle bearing is required bi-annually. Use only BioOps certified synthetic lubricants." }